SHDA100 SERIES |
|
100W Desktop Power Supply for Medical Equipment |
|
|
|
|
|
2 Year Warranty |
|
Approvals: |
|
Single Output |
||||
|
Product Number |
Output Voltage |
Max. Output Current |
Total Regulation |
Maximum Output Power |
|
SHDA100-S05 |
11 ~ 13 VDC |
9.09 ~ 7.69 A |
5% |
100W |
|
SHDA100-S06 |
13 ~ 16 VDC |
7.69 ~ 6.25 A |
4% |
100W |
|
SHDA100-S07 |
16 ~ 21 VDC |
6.25 ~ 4.76 A |
4% |
100W |
|
SHDA100-S08 |
21 ~ 27 VDC |
4.76 ~ 3.70 A |
4% |
100W |
|
SHDA100-S09 |
27 ~33 VDC |
3.70 ~ 3.03 A |
3% |
100W |
|
SHDA100-S11 |
40 ~ 48 VDC |
2.50 ~ 2.08 A |
3% |
100W |
|
Total Regulation is guaranteed by below configuration (S05: AWG16/2C/4FT output cable)(S06, S07: AWG18/2C/4FT output cable)(S08~S10: AWG18/2C/6FT output cable) |
||||
|
Electrical Characteristics |
|||||
|
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
|
Input Voltage |
Operating Voltage |
90 |
|
260 |
VAC |
|
Input Frequency |
|
47 |
|
63 |
Hz |
|
Power Factor Correction |
Io=Full load, Vin=230 VAC |
0.95 |
|
1 |
|
|
Output Power Range |
Vin= 90 to 264 VAC |
0 |
|
100 |
W |
|
Input Current (Low Line) |
Io=Full load, Vin=115 VAC |
|
|
1.35 |
A |
|
Input Current (High Line) |
Io=Full load, Vin= 230 VAC |
|
|
0.5 |
A |
|
Low Line Inrush Current |
Io=Full load, 25°C, Cool start, Vin=115VAC |
|
44 |
50 |
A |
|
High Line Inrush Current |
Io=Full load, 25°C, Cool start, Vin=230 VAC |
|
85 |
100 |
A |
|
Efficiency |
Io=Full Load, Vin=230VAC |
85 |
87 |
90 |
% |
|
Line Regulation |
Io=Full Load |
|
0.5 |
1 |
% |
|
Load Regulation |
Vin=230VAC |
|
3 |
5 |
% |
|
Over Voltage Protection |
|
112 |
|
132 |
% |
|
Over Current Protection |
|
110 |
|
150 |
% |
|
Transient Response |
Io=Full Load to Half Load, Vin=100VAC |
|
|
4 |
mS |
|
Hold-Up Time |
Io=Full Load, Vin=110VAC |
12 |
|
|
mS |
|
Start Up Time |
Io=Full Load, Vin=100VAC |
0.3 |
1.5 |
2 |
S |
|
* Ripple & Noise (Peak to Peak) |
Full Load, Vin=90VAC |
|
0.5 |
1 |
% |
|
Safety Ground Leakage Current |
Io= Full Load, Vin=240VAC |
|
0.15 |
0.25 |
mA |
|
No-Load Power Consumption |
No load, Vin=240VAC |
|
|
0.5 |
W |
|
Temperature Coefficient |
All output |
-0.04 |
|
0.04 |
%/°C |
|
Conditions |
|||||
|
Parameter |
Test Conditions |
Min. |
Typ. |
Max. |
Unit |
|
Operating Temperature |
|
0 |
|
70 |
°C |
|
Storage Temperature |
|
-40 |
|
85 |
°C |
|
Relative Humidity |
|
5 |
|
95 |
% |
|
Operation temperature at 25??/span>, calculated per MIL-HDBK-217F |
0.1 |
|
|
MHrs |
|
|
Derate linearly from 100% load at 50°C to 50% load at 70°C |
|||||
|
Approvals and Compliances |
|||
|
Parameter |
Test Conditions |
Min. |
Unit |
|
Dielectric Withstanding Voltage for Primary to secondary |
Primary to secondary |
5656 |
VDC |
|
Dielectric Withstanding Voltage for Primary to Ground |
Primary to ground |
2828 |
VDC |
|
Isolation Resistance |
Test Voltage = 500VDC |
50 |
MΩ |
|
EMI requirements for CISPR-11 |
Vin=220VAC |
B |
CLASS |
|
EMI requirements for FCC PART-18 |
Vin=110VAC |
B |
CLASS |
|
Safety UL/c-UL, TUV/T-mark, FCC, CE, CB |
UL 60601-1, EN 60601-1, IEC 60601-1 |
n/a |
n/a |
|
Environmental Compliance |
RoHS, Energy Star 2.0, CEC Level V |
|
|
|
Mechanical and PIN out |
|
|
Note:
|